Arq. Bras. Cardiol. 2025; 122(3): e20240450

Subclinical Atrial Fibrillation Screening in Dialytic Chronic Kidney Disease Patients Using Portable Device

Adson Patrik Vieira Carvalho ORCID logo , Gabriel Assis Lopes do Carmo ORCID logo , Cassia Aparecida Silva, Ana Cecília Oliveira ORCID logo , Lucas Giandoni Perez ORCID logo , Lilian Pires de Freitas do Carmo ORCID logo , Antonio L. Ribeiro ORCID logo

DOI: 10.36660/abc.20240450i

This Original Article is referred by the Short Editorial "Moving from the “Check the Pulse” Era To “Check the ECG” in the Prevention of Thromboembolic Events in Atrial Fibrillation".

Abstract

Background

Cardiovascular morbidity and mortality rates are higher in hemodialysis (HD) patients, with an increased prevalence of arrhythmias. Atrial fibrillation (AF) is an independent risk factor for mortality and thromboembolic events in dialysis patients. For a better understanding and management of AF in these patients, it is important to know its prevalence. The use of a portable device would be pioneering for this group of patients.

Objective

To screen HD patients for AF using a portable gadget and evaluate the device’s diagnostic performance.

Methods

HD patients at a tertiary hospital underwent AF screening during HD sessions using MyDiagnostick® (Applied Biomedical Systems). Multiple data were collected to evaluate potential associations. Statistical significance was defined as p < 0.05.

Results

388 patients were evaluated (female, 40.7%; mean age of 56.8 years old, SD ± 14.7; and HD time of 27 months, 10-57). Screening was positive in 16 (4.1%) patients. AF was confirmed by electrocardiogram in 7 (1.8%) patients. Male sex (p = 0.019), older age (p = 0.007), altered baseline electrocardiogram (p < 0.001), increased serum potassium (p = 0.021), reduced systolic blood pressure at the beginning of dialysis (p = 0.007), and stable angina (0.011) were associated with positive screening for AF. The device presented a 91.74% specificity (95% CI, 86.65% to 96.91%) and 100% sensitivity (95% CI, 100% to 100%), with a negative predictive value of 100% (95% CI, 100% to 100%) for AF screening.

Conclusion

The use of this device proved to be practical, with high sensitivity and excellent negative predictive value. Subclinical AF has a high prevalence and may be underestimated in this population.

Subclinical Atrial Fibrillation Screening in Dialytic Chronic Kidney Disease Patients Using Portable Device